A selective, accurate, and precise stability-indicating HPLC method for analysis of levamisole hydrochloride in bulk and in their injection and oral solution formulations has been developed and validated in accordance with ICH guidelines. The chromatographic separation was carried out on a C8 column (250 × 4.6mm with a particle size of 5 micrometer) using a mixture of phosphate buffer pH 8.0 and acetonitrile (70 : 30 v/v) as the mobile phase pumped at a flow rate of 1.5 ml/min with UV detection at 215 nm. The calibration curve was linear over the 10–50 μg/ml concentration range with a correlation coefficient of 1.0000. The limit of detection (LOD) and the limit of quantitation (LOQ) were 0.29 μg/ml and 0.89 μg/ml, respectively. The accuracy and the precision of the developed method were significantly good (RSD < 2%). The validity of the proposed method was further confirmed through the statistical comparison of the obtained data with those of the official method.