Octreotide is a synthetic cyclic octapeptide analogue of somatostatin-14. It is mainly administered for the treatment of acromegaly, severe diarrhea, and neuroendocrine neoplasias. In this work, a hydrophilic interaction liquid chromatography (HILIC) method with fluorescence (FL) detection was developed and validated for the quantitation of octreotide in solutions for injection. Chromatographic separation was performed on an XBridge®-HILIC analytical column under isocratic elution with a short chromatographic run time of less than 10 min. The mobile phase consisted of ammonium bicarbonate 8.6 mM (pH 8.1)/acetonitrile 35/65 (v/v). The high sensitivity and selectivity of the fluorescence detection, with the excitation wavelength (λexcitation) set at 280 nm and the emission wavelength set at (λemission) 330 nm, enabled a simple sample preparation procedure that included only dilution steps. The calibration curve showed good linearity with a correlation coefficient greater than 0.998. The method was successfully applied to the analysis of commercially available octreotide injection forms.